Pacemaker Lead Placement via Transmural Approach in an Adult With Palliated Single Ventricle Heart Disease

Given the lack of systemic venous return to the heart, palliated single ventricle patients frequently require epicardial pacemaker implantation for management of dysrhythmias including sinus node dysfunction, atrial arrhythmias, and heart block. Repeated device hardware replacement, frequently requi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:World journal for pediatric & congenital heart surgery 2022-07, Vol.13 (4), p.514-516
Hauptverfasser: Tunks, Robert D., Myers, John L., Cohen, Mark H., Moser, Kevin, Imundo, Jason R.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Given the lack of systemic venous return to the heart, palliated single ventricle patients frequently require epicardial pacemaker implantation for management of dysrhythmias including sinus node dysfunction, atrial arrhythmias, and heart block. Repeated device hardware replacement, frequently required due to high lead thresholds or other device failure, is a challenging and significant problem for this population. 3-dimensional imaging can assist in delineating the cardiac anatomy allowing for novel approaches to intervention. We review a patient with extracardiac Fontan circulation who underwent placement of an endocardial atrial pacemaker lead via a transmural approach with a 3D-printed model used for procedural guidance.