SIR-Spheres activity measurements reveal systematic miscalibration
Accurate dosimetry-guided radiopharmaceutical therapy fundamentally relies on knowledge of the quantity of radioactivity administered to patients. The purpose of this work was to perform an independent and NIST-traceable activity measurement of Y SIR-Spheres®. Gamma spectroscopic measurements of the...
Gespeichert in:
Veröffentlicht in: | Journal of Nuclear Medicine 2022-08, Vol.63 (8), p.1131-1135 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Accurate dosimetry-guided radiopharmaceutical therapy fundamentally relies on knowledge of the quantity of radioactivity administered to patients. The purpose of this work was to perform an independent and NIST-traceable activity measurement of
Y SIR-Spheres®.
Gamma spectroscopic measurements of the
Y internal pair production decay mode were made using a high-purity germanium detector. Un-modified patient SIR-Spheres® vials were placed within a high-density polyethylene source holder positioned at a distance of 210 cm from the detector, with acquisition durations of 3 - 6 hours. Measured annihilation radiation detection rates were corrected for radioactive decay during acquisition, dead time, source attenuation, and source geometry effects. Detection efficiency was determined by two independent and NIST-traceable methods. Resulting
Y activity measurements were compared against the manufacturer activity calibration.
Measured SIR-Sphere® vials (
= 5) were found to contain more activity than specified by the manufacturer calibration - on average the ratio of measured activity to calibrated was 1.233 ± 0.030. Activity measurements made using two distinct efficiency calibration methods were found to agree within 1%. Uncertainty in individual measurements was dominated by counting statistical uncertainty (~2.5%), uncertainty in the internal pair production branching ratio of
Y (1.5%), and efficiency calibration (1.2% - 1.9%).
The primary SIR-Spheres® activity calibration appears to be a significant underestimate of true activity. This mis-calibration has likely been consistent for as long as the SIR-Sphere® product has been available. This finding should be independently verified, and steps should be taken by the manufacturer to establish an accurate and traceable activity standard. |
---|---|
ISSN: | 0161-5505 1535-5667 2159-662X |
DOI: | 10.2967/jnumed.121.262650 |