The effect of simple heat treatment on apatite formation on grit‐blasted/acid‐etched dental Ti implants already in clinical use

Grit‐blasted/acid‐etched titanium dental implants have a moderately roughened surface that is suitable for cell adhesion and exhibits faster osseointegration. However, the roughened surface does not always maintain stable fixation over a long period. In this study, a simple heat treatment at 600°C w...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biomedical materials research. Part B, Applied biomaterials Applied biomaterials, 2022-02, Vol.110 (2), p.392-402
Hauptverfasser: Ogura, Ayano, Yamaguchi, Seiji, Le, Phuc Thi Minh, Yamamoto, Kayoko, Omori, Michi, Inoue, Kazuya, Kato‐Kogoe, Nahoko, Nakajima, Yoichiro, Nakano, Hiroyuki, Ueno, Takaaki, Yamada, Tomohiro, Mori, Yoshihide
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Grit‐blasted/acid‐etched titanium dental implants have a moderately roughened surface that is suitable for cell adhesion and exhibits faster osseointegration. However, the roughened surface does not always maintain stable fixation over a long period. In this study, a simple heat treatment at 600°C was performed on a commercially available dental Ti implant with grit‐blasting/acid‐etching, and its effect on mineralization capacity was assessed by examining apatite formation in a simulated body fluid (SBF). The as‐purchased implant displayed a moderately roughened surface at the micrometer scale. Its surface was composed of titanium hydride accompanied by a small amount of alumina particles derived from the grit‐blasting. Heat treatment transformed the titanium hydride into rutile without evidently changing the surface morphology. The immersion in SBF revealed that apatite formed on the heated implant at 7 days. Furthermore, apatite formed on the Ti rod surface within 1 day when the metal was subjected to acid and heat treatment without blasting. These indicate that apatite formation was conferred on the commercially available dental implant by simple heat treatment, although its induction period was slightly affected by alumina particles remaining on the implant surface. The heat‐treated implant should achieve stronger and more stable bone bonding due to its apatite formation.
ISSN:1552-4973
1552-4981
DOI:10.1002/jbm.b.34915