Diagnostic reference level for adult pelvic examination in several hospitals of Taif and Kharaj city, Saudi Arabia
The current study was aimed to estimate the entrance surface air kerma (ESAK, mGy) for adult patients undergoing conventional radiography of Anteroposterior pelvis examination, and finally to establish a local diagnostic reference level (DRL). A total of 500 patients were exposed to diagnostic radia...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2022-02, Vol.180, p.110049-110049, Article 110049 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The current study was aimed to estimate the entrance surface air kerma (ESAK, mGy) for adult patients undergoing conventional radiography of Anteroposterior pelvis examination, and finally to establish a local diagnostic reference level (DRL). A total of 500 patients were exposed to diagnostic radiation in four hospitals (coded A, B, C, and D) in Taif and Kharaj city, Saudi Arabia, with different X-ray equipment specifications. Patient demographic data like age (y), body mass index (BMI) in kg/m2 as well as exposure factors and X-ray tube output were recorded. ESAK (mGy) was first calculated using the exposure data and tube output values, then the ESAK values were used to estimate entrance surface dose (ESD). The average BMI was 23.9 kg/m2. The mean tube potential used in A, B, C, and D hospitals and the corresponding estimated ESD were found to be 74.2, 69.8, 73, 76,7 kVp, and 2.54, 2.64, 2.94, 3.03 mGy respectively. The correlation coefficient between ESAK and BMI was found to be 0.98. When compared to computed radiography (CR), the conventional X-ray digital radiography reduces the radiation exposure in pelvic imaging by a factor of 1.18. The third quartile of median proposed a lower than the DRL of the previous studies.
•Entrance skin air kerma and dose were determined for patients underwent pelvis examination.•Correlation coefficient found to be 0.98 between ESAK and body mass index values.•Digital radiography reduces the radiation exposure by a factor of 1.18 than the CT.•Third quartile of ESAK values used to determine DRL and found to be lower than the literature. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2021.110049 |