Removal of "ruptured" pulmonary artery infusion port catheter by pigtail catheter combined with gooseneck trap: A case report

BACKGROUNDImplanted intravenous infusion port (IVAP) is indicated for patients undergoing chemotherapy, total parenteral nutrition and long-term antibiotic treatment. Among their complications, the rupture and migration of the catheter of an IVAP via internal jugular vein represents a very rare but...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:World journal of clinical cases 2021, Vol.9 (29), p.8820-8824
Hauptverfasser: Chen, Gui-Qin, Wu, Yang, Zhao, Kai-Fei, Shi, Rong-Shu
Format: Report
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:BACKGROUNDImplanted intravenous infusion port (IVAP) is indicated for patients undergoing chemotherapy, total parenteral nutrition and long-term antibiotic treatment. Among their complications, the rupture and migration of the catheter of an IVAP via internal jugular vein represents a very rare but potentially severe condition. CASE SUMMARYA 43-year-old woman was identified with a spontaneous fracture and migration of catheter of an IVAP via right internal jugular vein after adjuvant chemotherapy for left breast cancer. A computed tomography showed the fractured catheter of the IVAP in the pulmonary artery. Therefore, we conducted an emergency procedure to remove the catheter fragment by a pigtail catheter combined with a gooseneck trap. CONCLUSIONWhen the fractured catheter of an IVAP was detected, the special shape of the pigtail catheter in combination with the gooseneck trap successfully facilitated the removal of the dislodged catheter.
ISSN:2307-8960
2307-8960
DOI:10.12998/wjcc.v9.i29.8820