A heparin-functionalized covered stent prepared by plasma technology
In this study, the surface of the covered stent was treated by plasma technology to introduce amino functional groups, and glutaraldehyde and heparin were successfully grafted to prepare a heparin-functionalized covered stent (HPLCS). The preparation parameters such as plasma treatment power, plasma...
Gespeichert in:
Veröffentlicht in: | Journal of biomaterials applications 2022-02, Vol.36 (7), p.1243-1253 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this study, the surface of the covered stent was treated by plasma technology to introduce amino functional groups, and glutaraldehyde and heparin were successfully grafted to prepare a heparin-functionalized covered stent (HPLCS). The preparation parameters such as plasma treatment power, plasma treatment time, concentration of glutaraldehyde and heparin, and pH of heparin solution were studied in detail. The functionalized heparin covered stent can make the titer of heparin reach 1.23 ± 0.03 IU/cm2. In animal experiments, after implantation in pigs for 6 months, the titer of heparin can still reach 0.93 ± 0.05 IU/cm2. This work provides a good method for preparing heparin covered stent. |
---|---|
ISSN: | 0885-3282 1530-8022 |
DOI: | 10.1177/08853282211051871 |