Improving the detection limit in capillary isotachophoresis using asymmetric neutralisation reaction boundary
An online method involving transient electrokinetic dosing and ITP with neutralization reaction boundary (NRB) and/or carrier ampholyte‐free isoelectric focusing (CAF IEF) was developed for the preconcentration, preseparation, and analytical determination of glyphosate in aqueous samples containing...
Gespeichert in:
Veröffentlicht in: | Electrophoresis 2022-02, Vol.43 (3), p.417-424 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An online method involving transient electrokinetic dosing and ITP with neutralization reaction boundary (NRB) and/or carrier ampholyte‐free isoelectric focusing (CAF IEF) was developed for the preconcentration, preseparation, and analytical determination of glyphosate in aqueous samples containing low concentrations of the analyte of interest. Various parameters were investigated in the framework of an optimization study with the aim of achieving the maximum concentration limit of detection (cLOD) decrease in minimum time. The proposed method used CAF IEF and/or ITP with NRB. The sample was dosed to the column on the stationary reaction boundary (CAF IEF) and/or moving reaction boundary (ITP with NRB), whereat a sharp pH step exists. Here, charge reversal was due to the ampholytes, and/or acid accumulation occurred because of charge loss. Similarly, the accumulated sample was mobilized with TE and analyzed using classical ITP in the second analytical column. Glyphosate (GLY), the analyte of interest, was chosen as a model substance for ITP with NRB and preconcentration as well as focusing preconcentration and CAF IEF using the asymmetric purpose‐built NRB. On one side of the asymmetric boundary was the zone of acidic pH; while the opposite side comprised a neutral/basic non‐conductive zone of the ampholyte—in this case, GLY. Such an arrangement enables the use of a lower pH on the acidic side, which allows the focusing of strongly acidic ampholytes and the accumulation of weak acids. The electrolyte composition and the dosing time were optimized, and a 14‐fold accumulation was achieved in 25 min compared to that by classical ITP and a 180‐fold accumulation was achieved through CAF IEF and preconcentration with a glyphosate sample. Both methods are simple and can be conducted using all commercial ITP systems. |
---|---|
ISSN: | 0173-0835 1522-2683 |
DOI: | 10.1002/elps.202000398 |