CORRELATION BETWEEN ANTERIOR–POSTERIOR AND LATERAL DIMENSIONS WITH THE EFFECTIVE AND WATER-EQUIVALENT DIAMETERS IN AXIAL IMAGES FROM HEAD COMPUTED TOMOGRAPHY EXAMINATIONS

The study aims to correlate the effective diameter (Deff) and water-equivalent diameter (Dw) parameters with anterior–posterior (AP), lateral (LAT) and AP + LAT dimensions in order to estimate the patient dose in head CT examinations. Seventy-four patient datasets from head CT examinations were retr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation protection dosimetry 2021-11, Vol.196 (3-4), p.248-256
Hauptverfasser: Dewi, Winda Kusuma, Anam, Choirul, Hidayanto, Eko, Wati, Annisa Lidia, Dougherty, Geoff
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The study aims to correlate the effective diameter (Deff) and water-equivalent diameter (Dw) parameters with anterior–posterior (AP), lateral (LAT) and AP + LAT dimensions in order to estimate the patient dose in head CT examinations. Seventy-four patient datasets from head CT examinations were retrospectively collected. The patient’s sizes were calculated from the middle slice using a software of IndoseCT. Dw and Deff were plotted as functions of AP, LAT and AP + LAT dimensions. The best trendline fit for LAT and AP functions was a second order polynomial, which resulted in R2 of 0.89 for Deff vs LAT, 0.88 for Dw vs LAT, 0.92 for Deff vs AP and 0.91 for Dw vs AP. A linear correlation was found for Deff vs AP + LAT, Dw vs AP + LAT and Dw vs Deff with R2 of 0.97, 0.96 and 0.98, respectively.
ISSN:0144-8420
1742-3406