Inappropriate shock delivery by an implantable cardioverter-defibrillator due to electrical interference with a refrigerator in a 4-year-old child
We report a patient with long QT syndrome who received an inappropriate implantable cardioverter-defibrillator shock due to electrical interference from a refrigerator. This electrical interference was mistakenly detected as an episode of ventricular fibrillation and ended with an inappropriate deli...
Gespeichert in:
Veröffentlicht in: | Cardiology in the young 2022-04, Vol.32 (4), p.676-678 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report a patient with long QT syndrome who received an inappropriate implantable cardioverter-defibrillator shock due to electrical interference from a refrigerator. This electrical interference was mistakenly detected as an episode of ventricular fibrillation and ended with an inappropriate delivery of shock without any warning symptoms before. |
---|---|
ISSN: | 1047-9511 1467-1107 |
DOI: | 10.1017/S104795112100370X |