Comparison of selected serum biochemistry measurements between the Nova Prime Plus VET, Nova pHOx Ultra, and Beckman Coulter AU680 analyzers in dogs
Background Blood gas chemistry analyzers typically produce results faster and use smaller sample volumes than reference chemistry analyzers. However, results may not be comparable between blood gas chemistry analyzers and reference chemistry analyzers or between different models of blood gas chemist...
Gespeichert in:
Veröffentlicht in: | Veterinary clinical pathology 2021-09, Vol.50 (3), p.327-341 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
Blood gas chemistry analyzers typically produce results faster and use smaller sample volumes than reference chemistry analyzers. However, results may not be comparable between blood gas chemistry analyzers and reference chemistry analyzers or between different models of blood gas chemistry analyzers. This could suggest the use of separate reference intervals and, thus, has implications when making clinical decisions.
Objective
We aimed to perform method comparison studies to evaluate selected canine serum biochemical values obtained using the Nova Stat Profile Prime Plus VET (Prime Plus VET), Stat Profile Nova pHOx Ultra (Ultra), and Beckman Coulter AU680 (Beckman) analyzers. We hypothesized that the three analyzers would be identical within inherent imprecision.
Methods
Jugular venous blood samples were collected from 103 endurance‐trained sled dogs, and serum was harvested and stored for analysis. Results for serum chloride, potassium, sodium, creatinine, and urea nitrogen concentrations obtained from the Prime Plus VET and Ultra analyzers were compared with results from the Beckman analyzer, which was considered to be a reference method. Results for serum chloride, potassium, sodium, creatinine, urea nitrogen, and L‐lactate concentrations obtained from the Prime Plus VET and Ultra analyzers were compared. Passing‐Bablok regression and Bland‐Altman plots were used for method comparison.
Results
Significant (P |
---|---|
ISSN: | 0275-6382 1939-165X |
DOI: | 10.1111/vcp.13010 |