Tissues attached to retrieved leadless pacemakers: Histopathological evaluation of tissue composition in relation to implantation time and complications

Leadless pacemakers (LPs) have proven safe and effective, but device revisions remain necessary. Either replacing the LP or implanting a new adjacent LP is feasible. Replacement seems more appealing, but encapsulation and tissue adhesions may hamper the safety and efficacy of LP retrieval. We determ...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Heart rhythm 2021-12, Vol.18 (12), p.2101-2109
Hauptverfasser: Breeman, Karel T.N., du Long, Romy, Beurskens, Niek E.G., van der Wal, Allard C., Wilde, Arthur A.M., Tjong, Fleur V.Y., Knops, Reinoud E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Leadless pacemakers (LPs) have proven safe and effective, but device revisions remain necessary. Either replacing the LP or implanting a new adjacent LP is feasible. Replacement seems more appealing, but encapsulation and tissue adhesions may hamper the safety and efficacy of LP retrieval. We determined the incidence and cellular characteristics of tissue adherent to retrieved LPs and the potential implications for end-of-life strategy. All 15 consecutive successful Nanostim LP retrievals in a tertiary center were included. We assessed the histopathology of adherent tissue and obtained clinical characteristics. Adherent tissue was present in 14 of 15 retrievals (93%; median implantation duration 36 months; range 0–96 months). The tissue consisted of fibrosis (n = 2), fibrosis and thrombus (n = 9), or thrombus only (n = 3). In short-term retrievals (
ISSN:1547-5271
1556-3871
DOI:10.1016/j.hrthm.2021.08.025