Can the introduction of single-use flexible ureteroscopes increase the longevity of reusable flexible ureteroscopes at a high volume centre?
Objectives To assess whether the introduction of single use flexibles ureteroscopes (su-fURS) at our high-volume centre had an advantageous impact on the turn-over and breakage rates of reusable fURS (re-fURS). Methods We analysed re-fURS number of usages and breakages at our centre between February...
Gespeichert in:
Veröffentlicht in: | World journal of urology 2022, Vol.40 (1), p.251-256 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objectives
To assess whether the introduction of single use flexibles ureteroscopes (su-fURS) at our high-volume centre had an advantageous impact on the turn-over and breakage rates of reusable fURS (re-fURS).
Methods
We analysed re-fURS number of usages and breakages at our centre between February 2015 and December 2018. We recorded the number of usages for analysed scope between the first usage until a breakage requiring reconditioning. Usage count was restarted following each reconditioning episode. Since su-fURS (Lithovue, Boston Scientific, USA) were introduced at our center in September 2016, we had the chance to compare different re-fURS life cycles according to both su-fURS availability and usage intensity (i.e., number of su-fURS used during each re-fURS life cycle). We then explored the relationship between su-fURS usage intensity and reusable scope survival (i.e., number of utilizations before any breakage requiring reconditioning) using locally weighted scatterplot smoothing (LOWESS) approach.
Results
Five different re-fURSs were employed at our centre, for a total of 1820 usages and 40 breakages requiring reconditioning. The overall mean (SD) number of usages before breaking was 40 (22). After su-fURS introduction, mean (SD) re-fURS number of usages increased from 35 (22) to 49 (20), (+ 40%,
p
= 0.02). The relationship between su-fURS usage intensity and reusable scopes survival showed a linear survival increase after 10 or more su-fURS scopes were used per life cycle.
Conclusions
The life cycle of re-fURS increased by 40% after the introduction of su-fURS. Ten or more used su-fURS per life cycle were associated with increased re-fURS survival. |
---|---|
ISSN: | 0724-4983 1433-8726 |
DOI: | 10.1007/s00345-021-03808-0 |