Effect of the photon intensity selected at various positions under the photopeak on the mass attenuation coefficient of x-rays
The variation in the value of the mass attenuation coefficient of x‐rays in aluminium was studied by counting photons under various positions in the photopeak employing a proportional counter spectrometer system with a good geometrical set‐up. The effect of (small angle) scattering was also examined...
Gespeichert in:
Veröffentlicht in: | X-ray spectrometry 1992-05, Vol.21 (3), p.133-136 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The variation in the value of the mass attenuation coefficient of x‐rays in aluminium was studied by counting photons under various positions in the photopeak employing a proportional counter spectrometer system with a good geometrical set‐up. The effect of (small angle) scattering was also examined using three different sets of collimators of diameter 9, 6 and 3 mm and employing both a pure x‐ray source (55Fe) and a mixed radiation source (57Co). The μm values obtained at the FWHM position for each set of collimators agree closely with the theoretical values. |
---|---|
ISSN: | 0049-8246 1097-4539 |
DOI: | 10.1002/xrs.1300210307 |