Intracoronary Imaging of a Synergy Megatron Stent Fracture in Ostial Right Coronary Artery
The Synergy Megatron stent confers enhanced axial and radial forces, which makes it an ideal choice for ostial and calcified lesions. However, in this case, the stent fractured. The mechanical stress due to severe calcification at the ostium, followed by a swinging segment, probably caused hinge eff...
Gespeichert in:
Veröffentlicht in: | The Journal of invasive cardiology 2021-08, Vol.33 (8), p.E674 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Synergy Megatron stent confers enhanced axial and radial forces, which makes it an ideal choice for ostial and calcified lesions. However, in this case, the stent fractured. The mechanical stress due to severe calcification at the ostium, followed by a swinging segment, probably caused hinge effect and led to a stent fracture. The fluoroscopy and stent-enhancing techniques suggest and could confirm the diagnosis, but intracoronary imaging is mandatory for anatomy characterization in order to decide treatment and optimize the result. |
---|---|
ISSN: | 1557-2501 1557-2501 |
DOI: | 10.25270/jic/21.00090 |