Usefulness of the GenMark ePlex RPP assay for the detection of respiratory viruses compared to the FTD21 multiplex RT-PCR
Cartridge-based multiplex panels covering numerous pathogens offer an advantage of minimal hands-on-time and short time to result to commercial RT-PCR assays. In this study, we evaluated the performance of the ePlex respiratory pathogen panel (RPP) compared to the Fast Track Diagnostics Respiratory...
Gespeichert in:
Veröffentlicht in: | Diagnostic microbiology and infectious disease 2021-09, Vol.101 (1), p.115424-115424, Article 115424 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Cartridge-based multiplex panels covering numerous pathogens offer an advantage of minimal hands-on-time and short time to result to commercial RT-PCR assays. In this study, we evaluated the performance of the ePlex respiratory pathogen panel (RPP) compared to the Fast Track Diagnostics Respiratory pathogens 21 multiplex RT-PCR assay (FTD21) using 400 clinical respiratory samples. Discrepant results were further analysed by a reference nucleic acid amplification testing (NAT) and a composite reference approach was used for final interpretation. Discordant results were observed in 56 targets corresponding to 54 samples. Sensitivities and specificities were 85.5% and 99.9% for the ePlex RPP and 95.8% and 99.7% for the FTD21 system, respectively.
Altogether, the ePlex RPP is a valuable tool for the rapid detection of a number of different respiratory viruses with the exception of the coronavirus family (low sensitivity ranging from 50-80%) and samples with a low pathogen load (Ct values >33). |
---|---|