Diagnostic reference levels during fluoroscopically guided interventions using mobile C-arms in operating rooms: A national multicentric survey
•Diagnostic reference and achievable levels are proposed for 31 surgery procedures.•An important variability was observed between the surgical specialties.•The highest radiation exposure were found in cardiology and vascular specialties. To establish diagnostic reference levels (DRLs) and achievable...
Gespeichert in:
Veröffentlicht in: | Physica medica 2021-06, Vol.86, p.91-97 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Diagnostic reference and achievable levels are proposed for 31 surgery procedures.•An important variability was observed between the surgical specialties.•The highest radiation exposure were found in cardiology and vascular specialties.
To establish diagnostic reference levels (DRLs) and achievable levels (ALs) for the most common fluoroscopically guided interventions (FGIs) performed in operating rooms using mobile C-arm equipment.
A national survey was performed in 57 centers in France. Anonymous data from 6817 patients undergoing FGIs were prospectively collected over a period of 7 months. DRLs (third quartile of the distribution) and ALs (median of the distribution) were determined for each type of intervention in terms of kerma area product (KAP) and fluoroscopy time (FT).
DRLs and ALs were proposed for 31 procedure types related to seven surgical specialties: orthopedics (n = 9), urology (n = 3), vascular (n = 6), cardiology (n = 5), neurosurgery (n = 3), gastrointestinal (n = 3), and multi-specialty (n = 2). DRLs in terms of KAP ranged from 0.1 Gy·cm2 for hallux valgus to 78 Gy·cm2 for abdominal aortic aneurysm endovascular repair. A factor of 155 was obtained between the FTs for a herniated lumbar disk (0.2 min) and an abdominal aortic aneurysm endovascular repair (31 min). The highest variations were obtained within orthopedic procedures in terms of KAP (ratio 122) and within gastrointestinal procedures in terms of FT (ratio 9). Overall, the FGIs associated with the highest radiation exposure (KAP > 10 Gy·cm2) were found in the cardiology, vascular, and gastrointestinal specialties.
DRLs and ALs are suggested for a wide range of FGIs performed in operating rooms using a mobile C-arm. We aim at providing a practical optimization tool for medical physicists and surgeons. |
---|---|
ISSN: | 1120-1797 1724-191X |
DOI: | 10.1016/j.ejmp.2021.05.013 |