Validation of a 4D-MRI guided liver stereotactic body radiation therapy strategy for implementation on the MR-linac
. Accurate tumor localization for image-guided liver stereotactic body radiation therapy (SBRT) is challenging due to respiratory motion and poor tumor visibility on conventional x-ray based images. Novel integrated MRI and radiotherapy systems enable direct in-room tumor visualization, potentially...
Gespeichert in:
Veröffentlicht in: | Physics in medicine & biology 2021-05, Vol.66 (10), p.105010 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | . Accurate tumor localization for image-guided liver stereotactic body radiation therapy (SBRT) is challenging due to respiratory motion and poor tumor visibility on conventional x-ray based images. Novel integrated MRI and radiotherapy systems enable direct in-room tumor visualization, potentially increasing treatment accuracy. As these systems currently do not provide a 4D image-guided radiotherapy strategy, we developed a 4D-MRI guided liver SBRT workflow and validated all steps for implementation on the Unity MR-linac.
. The proposed workflow consists of five steps: (1) acquisition of a daily 4D-MRI scan, (2) 4D-MRI to mid-position planning-CT rigid tumor registration, (3) calculation of daily tumor midP misalignment, (4) plan adaptation using adapt-to-position (ATP) with segment-weights optimization and (5) adapted plan delivery. The workflow was first validated in a motion phantom, performing regular motion at different baselines (±5 to ±10 mm) and patient-derived respiratory signals with varying degrees of irregularity. 4D-MRI derived respiratory signals and 4D-MRI to planning CT registrations were compared to the phantom input, and gamma and dose-area-histogram analyses were performed on the delivered dose distributions on film. Additionally, 4D-MRI to CT registration performance was evaluated in patient images using the full-circle method (transitivity analysis). Plan adaption was further analyzed
by creating adapted treatment plans for 15 patients with oligometastatic liver disease.
. Phantom trajectories could be reliably extracted from 4D-MRI scans and 4D-MRI to CT registration showed submillimeter accuracy. The DAH-analysis demonstrated excellent coverage of the dose evaluation structures GTV and GTV
. The median daily rigid 4D-MRI to midP-CT registration precision in patient images was |
---|---|
ISSN: | 0031-9155 1361-6560 |
DOI: | 10.1088/1361-6560/abfada |