Establishing the Biofidelity of a Multiphysics Finite Element Model of the Human Heart
Purpose Accelerating development of new therapeutic cardiac devices remains a clinical and technical priority. High-performance computing and the emergence of functional and complex in silico models of human anatomy can be an engine to accelerate the commercialization of innovative, safe, and effect...
Gespeichert in:
Veröffentlicht in: | Cardiovascular engineering and technology 2021-08, Vol.12 (4), p.387-397 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
Accelerating development of new therapeutic cardiac devices remains a clinical and technical priority. High-performance computing and the emergence of functional and complex
in silico
models of human anatomy can be an engine to accelerate the commercialization of innovative, safe, and effective devices.
Methods
An existing three-dimensional, nonlinear model of a human heart with flow boundary conditions was evaluated. Its muscular tissues were exercised using electrophysiological boundary conditions, creating a dynamic, electro-mechanical simulation of the kinetics of the human heart. Anatomic metrics were selected to characterize the functional biofidelity of the model based on their significance to the design of cardiac devices. The model output was queried through the cardiac cycle and compared to
in vivo
literature values.
Results
For the kinematics of mitral and aortic valves and curvature of coronary vessels, the model’s performance was at or above the 95th percentile range of the
in vivo
data from large patient cohorts. One exception was the kinematics of the tricuspid valve. The model’s mechanical use environment would subject devices to generally conservative use conditions.
Conclusions
This conservative simulated use environment for heart-based medical devices, and its judicious application in the evaluation of medical devices is justified, but careful interpretation of the results is encouraged. |
---|---|
ISSN: | 1869-408X 1869-4098 |
DOI: | 10.1007/s13239-021-00538-7 |