Characteristics of a bolus created using thermoplastic sheets for postmastectomy radiation therapy
This study applied a “shell bolus,” an immobilizing thermoplastic shell locally thickened with extra layers over the radiation target, during postmastectomy radiation therapy (PMRT). We performed ion chamber and film measurements for a solid water phantom for thermoplastic sheets and a gel bolus for...
Gespeichert in:
Veröffentlicht in: | Radiological physics and technology 2021-06, Vol.14 (2), p.179-185 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study applied a “shell bolus,” an immobilizing thermoplastic shell locally thickened with extra layers over the radiation target, during postmastectomy radiation therapy (PMRT). We performed ion chamber and film measurements for a solid water phantom for thermoplastic sheets and a gel bolus for dosimetric characterization using a 6-MV X-ray flattening-filter-free (FFF) beam. The air gaps between the body surface for the gel and shell bolus were measured using computed tomography (CT) images in patients who underwent PMRT. This included seven and 13 patients treated with the gel and shell boluses, respectively. A comparison of the dose differences between a 10-mm gel bolus and a 9.6-mm-thick thermoplastic sheet at the surface and 5 cm below the surface showed a 4.2% higher surface dose and 0.5% lower dose at 5-cm depth for the thermoplastic sheet compared to those for the gel bolus. The mean (
p
= 0.029) and maximum (
p
|
---|---|
ISSN: | 1865-0333 1865-0341 |
DOI: | 10.1007/s12194-021-00618-2 |