Development and in vivo assessment of a novel MRI‐compatible headframe system for the ovine animal model
Background The brain of sheep has primarily been used in neuroscience as an animal model because of its similarity to the human brain, in particular if compared to other models such as the lissencephalic rodent brain. Their brain size also makes sheep an ideal model for the development of neurosurgi...
Gespeichert in:
Veröffentlicht in: | The international journal of medical robotics + computer assisted surgery 2021-08, Vol.17 (4), p.e2257-n/a |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
The brain of sheep has primarily been used in neuroscience as an animal model because of its similarity to the human brain, in particular if compared to other models such as the lissencephalic rodent brain. Their brain size also makes sheep an ideal model for the development of neurosurgical techniques using conventional clinical CT/MRI scanners and stereotactic systems for neurosurgery.
Methods
In this study, we present the design and validation of a new CT/MRI compatible head frame for the ovine model and software, with its assessment under two real clinical scenarios.
Results
Ex‐vivo and in vivo trial results report an average linear displacement of the ovine head frame during conventional surgical procedures of 0.81 mm for ex‐vivo trials and 0.68 mm for in vivo tests, respectively.
Conclusions
These trial results demonstrate the robustness of the head frame system and its suitability to be employed within a real clinical setting. |
---|---|
ISSN: | 1478-5951 1478-596X |
DOI: | 10.1002/rcs.2257 |