Unexplained syncope during pacemaker interrogation: Wand, AutoCapture, or lead—Which one is the elusive culprit?

A 65‐year‐old gentleman underwent dual chamber pacemaker implantation (DDDR, St Jude Medical) 7 years back for infra‐hisian complete heart block. He was completely asymptomatic and came for his annual routine check‐up. After undergoing ECG with and without magnet, he was prepared for device evaluati...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2021-07, Vol.44 (7), p.1231-1235
Hauptverfasser: Bera, Debabrata, Majumder, Suchit, Ghose, Subir, Kar, Ayan, Sarkar, Rakesh, Mukherjee, Sanjeev S.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A 65‐year‐old gentleman underwent dual chamber pacemaker implantation (DDDR, St Jude Medical) 7 years back for infra‐hisian complete heart block. He was completely asymptomatic and came for his annual routine check‐up. After undergoing ECG with and without magnet, he was prepared for device evaluation. After placing the programmer wand over the chest as soon as the “;interrogate” button on the programmer screen was pressed, the patient immediately experienced pre‐syncope but recovered instantly as the wand was promptly withdrawn. After taking him to the casualty room with all resuscitation measures in hand, a repeat attempt of interrogation was made after connecting ECG, which revealed reproducible loss of capture (LOC), exclusively during wand placement. A differential diagnosis of lead failure, battery depletion, or wand related issues were considered. However, serial ECGs recorded without wand raised the possibility of AutoCapture malfunction. With all precautions, the device was programmed to fixed ventricular output mode after which interrogation could be performed safely. There was a remaining battery longevity of 2 years with acceptable lead parameters and stable threshold. He continues to be asymptomatic at 10 months of follow up.
ISSN:0147-8389
1540-8159
DOI:10.1111/pace.14224