Single-pass 1-needle actuation versus single-pass 3-needle actuation technique for EUS-guided liver biopsy sampling: a randomized prospective trial (with video)
Several reports have validated EUS-guided liver biopsy sampling (EUS-LB) as safe and effective. Nineteen-gauge EUS aspiration (FNA) or core (fine-needle biopsy [FNB]) needles are used, but different needle techniques can yield variable outcomes. Some data show that 1 pass (single liver puncture) wit...
Gespeichert in:
Veröffentlicht in: | Gastrointestinal endoscopy 2021-09, Vol.94 (3), p.551-558 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Several reports have validated EUS-guided liver biopsy sampling (EUS-LB) as safe and effective. Nineteen-gauge EUS aspiration (FNA) or core (fine-needle biopsy [FNB]) needles are used, but different needle techniques can yield variable outcomes. Some data show that 1 pass (single liver puncture) with 1 actuation (1 to-and-fro needle movement) may be enough to obtain a satisfactory specimen. However, there has not been a head-to-head comparison of single versus multiple needle actuations for EUS-LB.
This was a prospective randomized trial of EUS-LB in 40 patients comparing tissue yields and adequacy using 1 pass, 1 actuation (1:1) versus 1 pass 3 actuations (1:3) of an FNB needle. The primary outcome was number of complete portal triads (CPTs). Secondary outcomes were length of the longest piece, aggregate specimen length, number of cores >9 mm, and adverse events (AEs). Computerized randomization determined selection (either 1:1 or 1:3 with fanning technique). Sample lengths were measured before pathologic processing.
Both groups had similar demographics and indications for EUS-LB. All biopsy samples were adequate for pathologic interpretation. Compared with 1:1, biopsy sampling with 1:3 yielded more CPTs (mean [standard deviation], 17.25 [6.2] vs 24.5 [9.88]; P < .008) and longer aggregate specimen length (6.89 cm [1.86] vs 12.85 cm [4.02]; P < .001). AEs were not statistically different between the techniques. No severe AEs were noted.
EUS-LB using the 1:3 technique produced longer liver cores with more CPTs than the 1:1 technique with an equivalent safety profile. Two needle passes are more likely to provide tissue adequacy according to the American Association for the Study of Liver Diseases guidelines. (Clinical trial registration number: UMIN 000040101.)