In vivo kinematical validated knee model for preclinical testing of total knee replacement
A computational knee model facilitates efficient component design evaluations and preclinical testing under various dynamic loadings. However, the development of a highly mimicked dynamic whole knee model with specified ligament constraints that provides high predictive accuracy with in-vivo experim...
Gespeichert in:
Veröffentlicht in: | Computers in biology and medicine 2021-05, Vol.132, p.104311-104311, Article 104311 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A computational knee model facilitates efficient component design evaluations and preclinical testing under various dynamic loadings. However, the development of a highly mimicked dynamic whole knee model with specified ligament constraints that provides high predictive accuracy with in-vivo experiments remains a challenge.
In the present study, a musculoskeletal integrated force-driven explicit finite-element knee model with tibiofemoral and patellofemoral joints constrained with detailed soft tissue was developed. A proportional-integral-derivative controller was concurrently added to the knee model to track the boundary conditions. The actuations of the quadriceps and hamstrings were predicted via a subject-specific musculoskeletal model and matched with electromyography results.
Compared to in-vivo fluoroscopic results in a gait cycle, the predicted results of the kinematics of the tibiofemoral joint exhibited an agreement in terms of tendency and magnitude (anterior–posterior translation: RMSE = 1.1 mm, r2 = 0.87; inferior–superior translation: RMSE = 0.83 mm, r2 = 0.84; medial–lateral translation: RMSE = 0.82 mm, r2 = 0.05; flexion–extension rotation: RMSE = 0.23°, r2 = 1; internal-external rotation: RMSE = 1.85°, r2 = 0.65; varus–valgus rotation: RMSE = 1.39°, r2 = 0.08). Contact mechanics, including the contact area, pressure, and stress, were synchronously simulated on the tibiofemoral and patellofemoral joints.
The study provides a calibrated knee model and a kinematical validation approach that can be widely used in preclinical testing and knee prosthesis design.
•A musculoskeletal integrated force-driven finite element knee model was developed and verified by experiments.•The predicted results of kinematics of the tibiofemoral joint exhibited an agreement with in-vivo fluoroscopic results.•Contact mechanics including the contact area, pressure, and stress can be synchronously simulated by the proposed knee model. |
---|---|
ISSN: | 0010-4825 1879-0534 |
DOI: | 10.1016/j.compbiomed.2021.104311 |