Basophil counting in hematology analyzers: time to discontinue?

Basophils (basophilic granulocytes) are the least abundant cells in blood. Nowadays, basophils are included in the complete blood count performed by hematology analyzers and therefore reported in practically all patients in whom hematologic investigations are requested. However, hematology analyzers...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical chemistry and laboratory medicine 2021-04, Vol.59 (5), p.813-820
1. Verfasser: Hoffmann, Johannes J. M. L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Basophils (basophilic granulocytes) are the least abundant cells in blood. Nowadays, basophils are included in the complete blood count performed by hematology analyzers and therefore reported in practically all patients in whom hematologic investigations are requested. However, hematology analyzers are not reliable enough to report clinically useful results. This is due to a combination of very high analytical imprecision and poor specificity, because the chemical and physical methods used for basophil counting in hematology analyzers are ill-defined and thus basophils are not well recognized by the analyzers. As a result, false basophil counts are quite common. In view of increasing analytical performance demands, hematology laboratories should stop reporting basophil counts produced by hematology analyzers. Suggestions for alternative pathways are presented for those situations where basophils are of clinical relevance.
ISSN:1434-6621
1437-4331
DOI:10.1515/cclm-2020-1528