In-line MRI-LINAC depth dose measurements using an in-house plastic scintillation dosimeter

Plastic scintillation dosimeters (PSDs) have many properties that make them desirable for relative dosimetry with MRI-LINACs. An in-house PSD, Farmer ionisation chamber and Gafchromic EBT3 film were used to measure central axis percentage depth dose distributions (PDDs) at the Australian MRI-LINAC....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical physics & engineering express 2021-03, Vol.7 (2), p.25012
Hauptverfasser: Madden, Levi, Roberts, Natalia, Jelen, Urszula, Dong, Bin, Holloway, Lois, Metcalfe, Peter, Rosenfeld, Anatoly, Li, Enbang
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Plastic scintillation dosimeters (PSDs) have many properties that make them desirable for relative dosimetry with MRI-LINACs. An in-house PSD, Farmer ionisation chamber and Gafchromic EBT3 film were used to measure central axis percentage depth dose distributions (PDDs) at the Australian MRI-LINAC. Mean errors were calculated between each detector's responses, where the in-house PSD was on average within 0.7 % of the Farmer chamber and 1.4 % of film, while the Farmer chamber and film were on average within 1.1 % of each other. However, the PSD systematically over-estimated the dose as depth increased, approaching a maximum overestimation of the order of 3.5 % for the smallest field size measured. This trend was statistically insignificant for all other field sizes measured; further investigation is required to determine the source of this effect. The calculated values of mean absolute error are comparable to the those of trusted dosimeters reported in the literature. These mean absolute errors, and the ubiquity of desirable dosimetric qualities to PSDs suggest that PSDs in general are accurate for relative dosimetry with the MRI-LINAC. However, care should be taken when applying PSDs with MRI-LINACs, given the reported systematic trends dependent on field-size and depth of measurement.
ISSN:2057-1976
2057-1976
DOI:10.1088/2057-1976/abe295