Dual‐storage phosphor proton therapy dosimetry: Simultaneous quantification of dose and linear energy transfer
Purpose To investigate the feasibility of using the high Zeff storage phosphor material BaFBrI:Eu2+ in conjunction with the low Zeff storage phosphor material KCl:Eu2+ for simultaneous proton dose and linear energy transfer (LET) measurements by (a) measuring the fundamental optical and dosimetric p...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2021-04, Vol.48 (4), p.1941-1955 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To investigate the feasibility of using the high Zeff storage phosphor material BaFBrI:Eu2+ in conjunction with the low Zeff storage phosphor material KCl:Eu2+ for simultaneous proton dose and linear energy transfer (LET) measurements by (a) measuring the fundamental optical and dosimetric properties of BaFBrI:Eu2+, (b) evaluating its compatibility in being readout simultaneously with KCl:Eu2+ dosimeters, and (c) modeling and validating its LET dependence under elevated proton LET irradiation.
Methods
A commercial BaFBrI:Eu2+ storage phosphor detector (Model ST‐VI, Fujifilm) was characterized with energy dispersive x‐ray spectroscopy (EDS) analysis to obtain its elemental composition. The dosimeters were irradiated using both a Mevion S250 proton therapy unit (at the center of a spread‐out Bragg peak, SOBP) and a Varian Clinac iX linear accelerator with the latter being a low LET irradiation. The photostimulated luminescence (PSL) emission spectra, excitation spectra, and luminescent lifetimes of the detector were measured after proton and photon irradiations. Dosimetric properties including dose linearity, dose rate dependence, radiation hardness, temporal, and readout stabilities were studied using a laboratory optical reader after proton irradiations. In addition, its proton energy dependence was analytically modeled and experimentally validated by irradiating the detectors at various depths within the SOBP (Range: 15.0 g/cm2, Modulation: 10.0 g/cm2).
Results
The active detector composition for the high Zeff storage phosphor detector was found to be BaFBr0.85I0.15:Eu2+. The BaFBr0.85I0.15:Eu2+ material’s excitation and emission spectra were in agreement under proton and photon irradiations, with peaks of 586 ± 1 nm and 400 ± 1 nm, respectively, with a full width at half maximum (FWHM) of 119 ± 3 nm and 30 ± 2 nm, respectively. As dosimeter response under photon irradiation is generally believed to be free from LET effect, these results suggest LET independence of charge storage center types resulted from ionizing radiations. There is sufficient spectral overlaps with KCl:Eu2+ dosimeters allowing both dosimeters to be readout under equivalent readout conditions, that is, 594 nm stimulation and 420 nm detection wavelengths. Its PSL characteristic lifetime was found to be less than 5 microseconds which would make it suitable for fast 2D readout post irradiation. Its 420 nm emission band intensity was found to be linear up to 10 Gy absolute proton d |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1002/mp.14748 |