Non-infective left ventricular lead complications requiring re-intervention following cardiac resynchronization therapy: prevalence, causes and outcomes
Purpose Left ventricular (LV) lead complications in cardiac resynchronization therapy are challenging and poorly reported. We aimed to establish prevalence, causes and outcomes of LV lead complications requiring re-intervention. Methods We analysed the rate of complications in 2551 consecutive patie...
Gespeichert in:
Veröffentlicht in: | Journal of interventional cardiac electrophysiology 2022, Vol.63 (1), p.69-75 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
Left ventricular (LV) lead complications in cardiac resynchronization therapy are challenging and poorly reported. We aimed to establish prevalence, causes and outcomes of LV lead complications requiring re-intervention.
Methods
We analysed the rate of complications in 2551 consecutive patients who received a transvenous
de novo
LV lead as part of a cardiac resynchronization therapy device between 2000 and 2018. LV lead complications requiring re-intervention were identified; those due to infection were excluded. Patient, procedural and device characteristics, and outcomes were examined for non-infective LV lead complications requiring re-intervention.
Results
During a median of 4.7 years, 142 (5.6%) patients required re-intervention for non-infective LV lead complications with a decrease from 10.7% between 2000 and 2004, 8.7% between 2005 and 2009, 3.2% between 2010 and 2014 to 3.2% after 2014. The most common complications were LV lead displacement (50%), high pacing threshold (28%) and phrenic nerve stimulation (15%). Of the complications, 79 (56%) occurred within 90 days post-implant and 63 (44%) later. At the end of the study period, 132/142 patients (93%) had a functional LV lead. Lead re-intervention was associated with higher risk of complications (20%), but no increase in mortality (
P
= 0.19). Quadripolar leads had longer longevity and lower risk of complications compared with unipolar and bipolar LV leads.
Conclusions
A small but significant proportion of patients required LV lead re-intervention for complications following
de novo
implant. Lead displacement accounted for half of the re-interventions. Re-intervention was associated with a higher complication rate, but 92% of these patients had functional LV leads at the end of follow-up. |
---|---|
ISSN: | 1383-875X 1572-8595 |