Long-Term Safety and Feasibility of Left Bundle Branch Pacing in a Large Single-Center Study

Left bundle branch pacing (LBBP) is a novel pacing method and has been observed to have low and stable pacing thresholds in prior small short-term studies. The objective of this study was to evaluate the feasibility and safety of LBBP in a large consecutive diverse group of patients with long-term f...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Circulation. Arrhythmia and electrophysiology 2021-02, Vol.14 (2), p.e009261-e009261
Hauptverfasser: Su, Lan, Wang, Songjie, Wu, Shengjie, Xu, Lei, Huang, Zhouqing, Chen, Xiao, Zheng, Rujie, Jiang, Limeng, Ellenbogen, Kenneth A., Whinnett, Zachary I., Huang, Weijian
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Left bundle branch pacing (LBBP) is a novel pacing method and has been observed to have low and stable pacing thresholds in prior small short-term studies. The objective of this study was to evaluate the feasibility and safety of LBBP in a large consecutive diverse group of patients with long-term follow-up. This study prospectively enrolled 632 consecutive pacemaker patients with attempted LBBP from April 2017 to July 2019. Pacing parameters, complications, ECG, and echocardiographic measurements were assessed at implant and during follow-up of 1, 6, 12, and 24 months. LBBP was successful in 618/632 (97.8%) patients according to strict criteria for LBB capture. Mean follow-up time was 18.6±6.7 months. Two hundred thirty-one patients had follow-up over 2 years. LBB capture threshold at implant was 0.65±0.27 mV at 0.5 ms and 0.69±0.24 mV at 0.5 ms at 2-year follow-up. A significant decrease in QRS duration was observed in patients with left bundle branch block (167.22±18.99 versus 124.02±24.15 ms,
ISSN:1941-3084
1941-3149
1941-3084
DOI:10.1161/CIRCEP.120.009261