Intermediate-term performance and safety of His-bundle pacing leads: A single-center experience
The short-term safety, feasibility, and performance of His-bundle pacing (HBP) leads have been reported; however, their longer-term performance beyond 1 year remains unclear. The purpose of this study was to examine the intermediate-term performance and safety of HBP. All HBP lead implants at Virgin...
Gespeichert in:
Veröffentlicht in: | Heart rhythm 2021-05, Vol.18 (5), p.743-749 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The short-term safety, feasibility, and performance of His-bundle pacing (HBP) leads have been reported; however, their longer-term performance beyond 1 year remains unclear.
The purpose of this study was to examine the intermediate-term performance and safety of HBP.
All HBP lead implants at Virginia Commonwealth University between January 2014 and January 2019 were analyzed. HBP was performed using a Medtronic SelectSecure 3830-69 cm pacing lead.
Of 295 attempts, successful HBP implantation (selective or nonselective) was seen in 274 cases (93%). Mean follow-up duration was 22.8 ± 19.5 months (median 19.5; interquartile range 11–33). Mean age was 69 ± 15 years; 58% were males; and ejection fraction |
---|---|