Variations in non-locking screw insertion conditions generate unpredictable changes to achieved fixation tightness and stripping rates
Screws are the most commonly inserted orthopaedic implants. However, several variables related to screw insertion and tightening have not been evaluated. This study aimed firstly to assess the effect of insertion variables on screw tightness, secondly to improve methodologies used by researchers whe...
Gespeichert in:
Veröffentlicht in: | Clinical biomechanics (Bristol) 2020-12, Vol.80, p.105201-105201, Article 105201 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Screws are the most commonly inserted orthopaedic implants. However, several variables related to screw insertion and tightening have not been evaluated. This study aimed firstly to assess the effect of insertion variables on screw tightness, secondly to improve methodologies used by researchers when testing screw insertion techniques and thirdly to assess for any learning or fatigue effects when inserting screws.
Two surgeons tightened a total of 2280 non-locking, 3.5 mm cortical screws, with 120 screws inserted to what they felt to be optimum tightness whilst varying each of the following factors: different screwdrivers for measuring torque, screwdriver orientation, gloves usage, dominant/non-dominant hand usage, awareness to the applied torque (blinded, unblinded and re-blinded), four bone densities and seven cortical thicknesses. Screws were tightened to failure to determine stripping torque, which was used to calculate screw tightness – ratio between stopping and stripping torque.
Screw tightness increased with glove usage, being blinded to the applied torque and with denser artificial bone and with thinner cortices. Considering all the insertions performed, the two surgeons stopped tightening screws at difference values of tightness ((77% versus 66% (p |
---|---|
ISSN: | 0268-0033 1879-1271 |
DOI: | 10.1016/j.clinbiomech.2020.105201 |