Characterization of tissue properties in epidural needle insertion on human specimen and synthetic materials

The force experienced while inserting an 18-gauge Tuohy needle into the epidural space or dura is one of only two feedback components perceived by an anaesthesiologist to deduce the needle tip position in a patient's spine. To the best of the authors knowledge, no x-ray validated measurements o...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the mechanical behavior of biomedical materials 2020-10, Vol.110, p.103946-103946, Article 103946
Hauptverfasser: Esterer, Benjamin, Hollensteiner, Marianne, Schrempf, Andreas, Winkler, Martin, Gabauer, Stefan, Fürst, David, Merwa, Robert, Panzer, Stephanie, Püschel, Klaus, Augat, Peter
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The force experienced while inserting an 18-gauge Tuohy needle into the epidural space or dura is one of only two feedback components perceived by an anaesthesiologist to deduce the needle tip position in a patient's spine. To the best of the authors knowledge, no x-ray validated measurements of these forces are currently available to the public. A needle insertion force recording during an automated insertion of an 18-gauge Tuohy needle into human vertebral segments of four female donors was conducted. During the measurements, x-ray images were recorded simultaneously. The force peaks due to the penetration of the ligamentum supraspinale and ligamentum flavum were measured and compared to the measurements of an artificial patient phantom for a hybrid patient simulator. Based on these force peaks and the slope of the ligamentum interspinale, a mathematical model was developed. The model parameters were used to compare human specimens and artificial patient phantom haptics. The force peaks for the ligamenta supraspinale and flavum were 7.55 ± 3.63 N and 15.18 ± 5.71 N, respectively. No significant differences were found between the patient phantom and the human specimens for the force peaks and four of six physical model parameters. The patient phantom mimics the same resistive force against the insertion of an 18-gauge Tuohy needle. However, there was a highly significant (p 
ISSN:1751-6161
1878-0180
DOI:10.1016/j.jmbbm.2020.103946