Towards MR-guided electron therapy: Measurement and simulation of clinical electron beams in magnetic fields
•Therapeutic small electron beams are investigated in MRI-linac style magnetic fields.•A comprehensive Monte Carlo framework is established to support the experiments.•Focusing of electron beams & reduced penumbra is observed in inline magnetic fields.•The results motivate investigation of an el...
Gespeichert in:
Veröffentlicht in: | Physica medica 2020-10, Vol.78, p.83-92 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Therapeutic small electron beams are investigated in MRI-linac style magnetic fields.•A comprehensive Monte Carlo framework is established to support the experiments.•Focusing of electron beams & reduced penumbra is observed in inline magnetic fields.•The results motivate investigation of an electron mode in future inline MRI-linacs.
In the current era of MRI-linac radiotherapy, dose optimization with arbitrary dose distributions is a reality. For the first time, we present new and targeted experiments and modeling to aid in evaluating the potential dose improvements offered with an electron beam mode during MRI-linac radiotherapy.
Small collimated (1 cm diameter and 1.5 × 1.5 cm2 square) electron beams (6, 12 and 20 MeV) from a clinical linear accelerator (Varian Clinac 2100C) are incident perpendicular and parallel to the strong and localized magnetic fields (0–0.7 T) generated by a permanent magnet device. Gafchromic EBT3 film is placed inside a slab phantom to measure two-dimensional dose distributions. A benchmarked and comprehensive Monte Carlo model (Geant4) is established to directly compare with experiments.
With perpendicular fields a 5% narrowing of the beam FWHM and a 10 mm reduction in the 15% isodose penetration is seen for the 20 MeV beam. In the inline setup the penumbral width is reduced by up to 20%, and a local central dose enhancement of 100% is observed. Monte Carlo simulations are in agreement with the measured dose distributions (2% or 2 mm).
A new range of experiments have been performed to offer insight into how an electron beam mode could offer additional choices in MRI-linac radiotherapy. The work extends on historic studies to bring a successful unified experimental and Monte Carlo modeling approach for studying small field electron beam dosimetry inside magnetic fields. The results suggest further work, particularly on the inline magnetic field scenario. |
---|---|
ISSN: | 1120-1797 1724-191X |
DOI: | 10.1016/j.ejmp.2020.09.001 |