Intra- and inter-observer variability in intracranial aneurysm segmentation: comparison between CT angiography (semi-automated segmentation software stroke VCAR) and digital subtraction angiography (3D rotational angiography)

Purpose To compare size and morphologic features of three-dimensional aneurysm models, obtained with a semi-automated segmentation software (Stroke VCAR, GE, USA) from cerebral CT angiography (CTA) data, to three-dimensional aneurysm models obtained with digital subtraction angiography (DSA, with 3D...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiologia medica 2021-03, Vol.126 (3), p.484-493
Hauptverfasser: D’Argento, F., Pedicelli, A., Ciardi, C., Leone, E., Scarabello, M., Infante, A., Alexandre, A., Lozupone, E., Valente, I., Colosimo, C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose To compare size and morphologic features of three-dimensional aneurysm models, obtained with a semi-automated segmentation software (Stroke VCAR, GE, USA) from cerebral CT angiography (CTA) data, to three-dimensional aneurysm models obtained with digital subtraction angiography (DSA, with 3D rotational angiography acquisition—3DRA), considered as the reference standard. Methods In this retrospective study, we reviewed 132 patients, with a total number of 137 intracranial aneurysm, who underwent CTA and subsequent DSA examination, supplemented with 3DRA. We compared neck length, short axis and long axis measured on 3DRA model to the same variables measured on 3D-CTA model by two blinded readers and to the automatic software dimensions. Therefore, statistics analysis assessed intra-observer and inter-observer variability and differences between patients with or without subarachnoid hemorrhage (SAH). Results There were no significant differences in short-axis and long-axis measurements between 3D angiographic and 3D-CTA models, while comparison of neck lengths revealed a statistically significant difference, which tended to be greater for smaller neck lengths (partial volume effect and “kissing vessels” artifact). There were significant differences between manual and automatic data measured for the same three variables, and the presence of SAH did not affect aneurysm 3D reconstruction. Inter-observer agreement resulted moderate for neck length and substantial for short axis and long axis. Conclusion The examined 3D-CTA segmentation system is a reproducible procedure for aneurysm morphologic characterization and, in particular, for assessment of aneurysm sac dimensions, but considerable carefulness is required in neck length interpretation.
ISSN:0033-8362
1826-6983
DOI:10.1007/s11547-020-01275-y