Accuracy of acetabular cup insertion in an anterolateral supine approach using an accelerometer-based portable navigation system
The purpose of this study was to determine the accuracy of acetabular cup insertion during total hip arthroplasty (THA) in a supine position using an accelerometer-based portable navigation system. A single-surgeon study was conducted in which 62 prospective patients with navigation and 42 retrospec...
Gespeichert in:
Veröffentlicht in: | Journal of artificial organs 2021-03, Vol.24 (1), p.82-89 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study was to determine the accuracy of acetabular cup insertion during total hip arthroplasty (THA) in a supine position using an accelerometer-based portable navigation system. A single-surgeon study was conducted in which 62 prospective patients with navigation and 42 retrospective patients without navigation as historical controls were compared. The patients underwent THA via an anterolateral supine approach. The absolute values of errors of radiographic inclination and anteversion were calculated. The navigation error was also calculated. The factors that affected the absolute values of errors of cup alignment were determined. The mean absolute error of the postoperative CT measurement from the target angle for radiographic inclination was 3.8° in the navigation group and 6.6° in the control group (
P
|
---|---|
ISSN: | 1434-7229 1619-0904 |
DOI: | 10.1007/s10047-020-01206-8 |