Application of Molecularly Imprinted Polymer-modified Potentiometric Sensor for Quantitative Determination of Histamine in Serum
A molecularly imprinted polymer-modified potentiometric histamine (HIS) sensor was prepared and used for quantitative determination of HIS in bovine serum. The calibration curve using the potential responses measured in 1 × 10−3 mol L−1 phosphate buffer (pH 7.4) showed good linearity in the HIS conc...
Gespeichert in:
Veröffentlicht in: | Analytical Sciences 2020/12/10, Vol.36(12), pp.1561-1563 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A molecularly imprinted polymer-modified potentiometric histamine (HIS) sensor was prepared and used for quantitative determination of HIS in bovine serum. The calibration curve using the potential responses measured in 1 × 10−3 mol L−1 phosphate buffer (pH 7.4) showed good linearity in the HIS concentration range of 3 × 10−4 to 1 × 10−2 mol L−1 (r = 0.92), with a detection limit of 1.6 × 10−4 mol L−1. In bovine serum samples, the HIS sensor showed good recovery values of 91 – 104%. Therefore, this HIS sensor successfully determined the HIS concentration in bovine serum samples. |
---|---|
ISSN: | 0910-6340 1348-2246 |
DOI: | 10.2116/analsci.20N018 |