The role of multiple observations in small‐molecule single‐crystal service X‐ray structure determination
In order to gain a better understanding of how to improve the quality of small‐molecule single‐crystal X‐ray diffraction data achievable in a finite time, a study was carried out to investigate the effect of varying the multiplicity, acquisition time, detector binning, maximum resolution and complet...
Gespeichert in:
Veröffentlicht in: | Acta crystallographica Section B, Structural science, crystal engineering and materials Structural science, crystal engineering and materials, 2019-08, Vol.75 (4), p.657-673 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In order to gain a better understanding of how to improve the quality of small‐molecule single‐crystal X‐ray diffraction data achievable in a finite time, a study was carried out to investigate the effect of varying the multiplicity, acquisition time, detector binning, maximum resolution and completeness. The results suggest that, unless there are strong arguments for a different strategy, a good routine procedure might be to optimize the conditions necessary to get the best data from single scans, and then choose a multiplicity of observations (MoO) to utilize the available time fully. Different strategies may be required if the crystal is highly absorbing, is larger than the incident beam, is enclosed in a capillary tube or is unusual in some other way. The signal‐to‐noise ratio should be used with care, as collecting data for longer or at higher multiplicity appears to give a systematic underestimate of the intensity uncertainties. Further, the results demonstrate that including poor‐quality data in a refinement may degrade the result and, in the general case, the accidental omission of reflections has a very small impact on the refinement as long as they are omitted at random. Systematic omission of reflections needs a convincing procedural justification.
The effect of varying the multiplicity of observations is explored in the context of varying acquisition time, detector binning, maximum resolution and completeness. The cost/benefit analysis suggests that, in the context of service analytical crystallography, seeking a high multiplicity of observations and including poor‐quality data in a refinement do not significantly improve the result. |
---|---|
ISSN: | 2052-5206 2052-5192 2052-5206 |
DOI: | 10.1107/S2052520619006681 |