55 Mn-based fiducial markers for rapid and automated RF coil localization for hyperpolarized 13 C MRI

To use fiducial markers containing manganese 55 to rapidly localize carbon 13 ( C) RF coils for correcting images for B variation. Hollow high-density polyethylene spheres were filled with 3M sodium permanganate and affixed to a rectangular C-tuned RF coil. The relative positions of the markers and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 2021-01, Vol.85 (1), p.518-530
Hauptverfasser: Ohliger, Michael A, Gordon, Jeremy W, Carvajal, Lucas, Larson, Peder E Z, Ou, Jao J, Agarwal, Shubhangi, Zhu, Zihan, Vigneron, Daniel B, von Morze, Cornelius
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To use fiducial markers containing manganese 55 to rapidly localize carbon 13 ( C) RF coils for correcting images for B variation. Hollow high-density polyethylene spheres were filled with 3M sodium permanganate and affixed to a rectangular C-tuned RF coil. The relative positions of the markers and coil conductors were mapped using CT. Marker positions were measured by MRI using a series of 1D projections and automated peak detection. Once the coil location was determined, coil sensitivity was estimated using a quasi-static calculation. Simulations were performed to determine the minimum number of projections required for robust localization. Phantom experiments were used to confirm the accuracy of marker localization as well as the calculated coil sensitivity. Finally, in vivo validation was performed using hyperpolarized C pyruvate in a rat model. In simulations, our algorithm was accurate in determining marker positions when at least 6 projections were used (RMSE 1.4 ± 0.9 mm). These estimates were verified in phantom experiments, where markers locations were determined with an RMS accuracy of 1.3 mm. A minimum SNR of 4 was required for automated detection to perform accurately. Computed coil sensitivity had a median error of 17% when taken over the entire measured area and 5.7% over a central region. In a rat, correction for nonuniform reception and flip angle was able to normalize the signals arising from asymmetrically positioned kidneys. Manganese 55 fiducial markers are an inexpensive and reliable method for rapidly localizing C RF coils and correcting C images for B variation without user intervention.
ISSN:0740-3194
1522-2594
DOI:10.1002/mrm.28424