Noninvasive measurement of arterial blood pressure in patients with continuous-flow left ventricular assist devices: a systematic review
Accurate mean blood pressure determination is essential to prevent adverse events in patients with continuous-flow left ventricular assist devices (CFLVAD). We sought to evaluate the accuracy of noninvasive methods of blood pressure measurement compared with invasive intra-arterial recordings in pat...
Gespeichert in:
Veröffentlicht in: | Heart failure reviews 2021, Vol.26 (1), p.47-55 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Accurate mean blood pressure determination is essential to prevent adverse events in patients with continuous-flow left ventricular assist devices (CFLVAD). We sought to evaluate the accuracy of noninvasive methods of blood pressure measurement compared with invasive intra-arterial recordings in patients with CFLVAD. Systematic electronic search was performed on four online databases (PubMed, Scopus, Embase, and Web of Knowledge) for the terms “Blood Pressure” AND (“Heart-Assist Devices” OR “Left ventricular Assist Devices”). Only studies that compared an intra-arterial and noninvasive blood pressure measurement were included. Electronic search of scientific literature identified 5968 articles. After deduplication, screening of titles and abstracts, full-text review, and excluding incorrect populations and comparator, a total of 12 studies with 502 participants were included, of those 402 participants who had intra-arterial blood pressure measurement. Doppler mean arterial blood pressure showed a very high correlation with mean intra-arterial blood pressure (
r
= 0.97,
r
= 0.87) in low pulsatility situations. When the pulsatility was not evaluated, the correlation was high moderate (
r
= 0.63,
r
= 0.741). In low pulsatility situations, the correlation was moderate to high moderate (
r
= 0.42 to
r
= 0.65). Oscillometer automatic blood pressure cuff showed a moderate to very high correlation with intra-arterial mean arterial blood pressure (
r
= 0.42,
r
= 0.86) but also could be low in the context of low pulsatility associated with inconsistent success in noninvasive measurement (
r
= 0.25). Studies correlating intra-arterial with noninvasive techniques were performed in the context of routine clinical care using fluid-filled catheters. The degree of correlation between both methods is at least moderate. |
---|---|