Sterility and performance of open and closed extracorporeal circuits after long-term dry-wet setups

Background: The timeframe for safely using previously setup dry, crystalloid, and blood-primed extracorporeal circuits has long been debated. This study was undertaken to determine a safe deviation from standardized recommendations. Methods: Open (cardiopulmonary bypass) circuits and closed extracor...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Perfusion 2021-03, Vol.36 (2), p.130-137
Hauptverfasser: Deptula, Joseph, McGrath, Catrina, Preston, Thomas, Miller, Hayden, Yen, Bianca, Munari, Lenny, Setty, Shaun P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background: The timeframe for safely using previously setup dry, crystalloid, and blood-primed extracorporeal circuits has long been debated. This study was undertaken to determine a safe deviation from standardized recommendations. Methods: Open (cardiopulmonary bypass) circuits and closed extracorporeal membrane oxygenation circuits were setup dry for up to 60 days and wet primed for up to 6 weeks with one control inoculated with Escherichia coli. Open circuits were cultured daily, closed circuits weekly. Circuits were primed with blood, albumin, heparin, NaHCO3, and CaCl2. Baseline pCO2, pO2, hemoglobin, lactate dehydrogenase, and plasma free hemoglobin were measured. Circuits were recirculated at a blood flow of 6 Liters/minute with a sweep gas of 1 Liter/minute at 100% FiO2 for 1 minute. Post oxygenator blood gases were collected at 8-, 16-, and 24-hour intervals. Results: There was no observed compromise to the sterility of the circuits and no clinically significant gas exchange abnormalities observed over the duration of the study period. Statistical significance (p 
ISSN:0267-6591
1477-111X
DOI:10.1177/0267659120937898