Scatter Radiation Intensities during Transforaminal Lumbar Interbody Fusion Using a Mobile C-arm System
The purpose of this study was to measure the scatter radiation intensity during transforaminal lumbar interbody fusion using a mobile C-arm system (Arcadis Orbic 3D; Siemens) and minimize radiation exposure. Dosimetry was performed with anterior-posterior and lateral continuous fluoroscopy, and cone...
Gespeichert in:
Veröffentlicht in: | Japanese Journal of Radiological Technology 2020, Vol.76(6), pp.572-578 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng ; jpn |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study was to measure the scatter radiation intensity during transforaminal lumbar interbody fusion using a mobile C-arm system (Arcadis Orbic 3D; Siemens) and minimize radiation exposure. Dosimetry was performed with anterior-posterior and lateral continuous fluoroscopy, and cone beam computed tomography (CT). A scaffold tower (L: 300 cm×W: 200 cm×H: 150 cm) was built with radiation-resistant paper cylinders at intervals of 50 cm and plastic joints over the bed, and 100 optically stimulated luminescence dosimeters (nanoDot; Nagase Landauer) were placed on each joint. A human torso phantom from head to pelvis (Kyoto Kagaku) was positioned on the bed in a prone position. The scatter radiation dose in a lateral view was highest on the X-ray tube side at the height of 100 cm (170.5 μGy/min). The scatter radiation dose increased significantly on the X-ray tube side during lateral continuous fluoroscopy. Continuous change of surgeons' standing positions is important to minimize radiation exposure received by a specific surgeon. |
---|---|
ISSN: | 0369-4305 1881-4883 |
DOI: | 10.6009/jjrt.2020_JSRT_76.6.572 |