Autopsy evaluation of the implantation site of a His bundle pacing lead demonstrating selective capture
Evaluations of His bundle pacing (HBP) lead location at autopsy examination have been rarely reported. We report an autopsy case of a 98‐year‐old man who underwent HBP implantation due to atrioventricular block and heart failure. Although selective HBP was achieved with an acceptable threshold, the...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 2020-11, Vol.43 (11), p.1412-1416 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Evaluations of His bundle pacing (HBP) lead location at autopsy examination have been rarely reported. We report an autopsy case of a 98‐year‐old man who underwent HBP implantation due to atrioventricular block and heart failure. Although selective HBP was achieved with an acceptable threshold, the stimulus‐to‐QRS interval was relatively longer without correction of the right bundle‐branch block. A macroscopic examination revealed that the HBP lead was inserted on the ventricular side passing through the anteroseptal commissure of the tricuspid valve. Transthyretin cardiac amyloidosis may affect the distal conduction system resulting in a long stimulus‐to‐QRS interval during selective HBP. |
---|---|
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/pace.13989 |