Discriminative Ability and Reliability of Transesophageal Echocardiography in Characterizing Cases of Cardiac Device Lead Vegetations Versus Noninfectious Echodensities
Abstract Background Approximately one-third of cases of cardiovascular implantable electronic device (CIED) infection present as CIED lead infection. The precise transesophageal echocardiographic (TEE) definition and characterization of “vegetation” associated with CIED lead infection remain unclear...
Gespeichert in:
Veröffentlicht in: | Clinical infectious diseases 2021-06, Vol.72 (11), p.1938-1943 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Background
Approximately one-third of cases of cardiovascular implantable electronic device (CIED) infection present as CIED lead infection. The precise transesophageal echocardiographic (TEE) definition and characterization of “vegetation” associated with CIED lead infection remain unclear.
Methods
We identified a sample of 25 consecutive cases of CIED lead infection managed at our institution between January 2010 and December 2017. Cases of CIED lead infection were classified using standardized definitions. Similarly, a sample of 25 noninfected patients who underwent TEE that showed a defined lead echodensity during the study period was included as a control group. TEEs were reviewed by 2 independent echocardiologists who were blinded to all linked patient demographic, clinical, and microbiological information. Reported echocardiographic variables of the infected vs noninfected cases were compared, and the overall diagnostic performance was analyzed.
Results
Descriptions of lead echodensities were variable and there were no significant differences in median echodensity diameter or mobility between infected vs noninfected groups. Among infected cases, blinded echocardiogram reports by either reviewer correctly made a prediction of infection in 6 of 25 (24%). Interechocardiologist agreement was 68%. Sensitivity of blinded TEEs ranged from 31.5% to 37.5%.
Conclusions
Infectious vs noninfectious lead echodensities could not be reliably distinguished on the basis of size, mobility, and general shape descriptors obtained from a retrospective blinded TEE examination without knowledge of clinical and microbiological parameters. Therefore, a reanalysis of criteria used to support a diagnosis of CIED lead infection may be warranted.
We demonstrate that infectious vs noninfectious cardiovascular implantable electronic device lead echodensities cannot be reliably distinguished on the basis of size, mobility, or general shape descriptors obtained from a retrospective blinded transesophageal echocardiographic examination without knowledge of clinical and microbiological parameters. |
---|---|