Low-temperature nuclear magnetic resonance with a dc SQUID amplifier

We report the detection of nuclear magnetic resonance at 4 K and 1.9 MHz with a receiver based on a thin-film dc superconducting quantum interference device. The noise temperature of the spectrometer is 300 mK, limited by the second stage amplifier.

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Appl. Phys. Lett.; (United States) 1986-01, Vol.48 (4), p.300-302
Hauptverfasser: FREEMAN, M. R, GERMAIN, R. S, RICHARDSON, R. C, ROUKES, M. L, GALLAGHER, W. J, KETCHEN, M. B
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report the detection of nuclear magnetic resonance at 4 K and 1.9 MHz with a receiver based on a thin-film dc superconducting quantum interference device. The noise temperature of the spectrometer is 300 mK, limited by the second stage amplifier.
ISSN:0003-6951
1077-3118
DOI:10.1063/1.96587