Cardiac Tissue-Mimicking Ballistic Gel Phantom for Ultrasound Imaging in Clinical and Research Applications

Ballistic gel was investigated as a tissue-mimicking material in an anthropomorphic cardiac phantom for ultrasound imaging. The gel was tested for its acoustic properties and its compatibility with conventional plastics molding techniques. Speed of sound and attenuation were evaluated in the range 2...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Ultrasound in medicine & biology 2020-08, Vol.46 (8), p.2057-2069
Hauptverfasser: Alves, Natasha, Kim, Angela, Tan, Jeremy, Hwang, Germain, Javed, Talha, Neagu, Bogdan, Courtney, Brian K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Ballistic gel was investigated as a tissue-mimicking material in an anthropomorphic cardiac phantom for ultrasound imaging. The gel was tested for its acoustic properties and its compatibility with conventional plastics molding techniques. Speed of sound and attenuation were evaluated in the range 2–12 MHz. The speed of sound was 1537 ± 39 m/s, close to typical values for cardiac tissue (∼1576 m/s). The attenuation coefficient was 1.07 dB/cm·MHz, within the range of values previously reported for cardiac tissue (0.81–1.81 dB/cm·MHz). A cardiac model based on human anatomy was developed using established image segmentation processes and conventional plastic molding techniques. Key anatomic features were observed, captured and identified in the model using an intracardiac ultrasound imaging system. These favorable results along with the material's durability and processes that allow for repetitive production of detailed whole-heart models at low cost are promising. There are numerous applications for geometrically complex phantoms in research, training, device development and clinical use.
ISSN:0301-5629
1879-291X
DOI:10.1016/j.ultrasmedbio.2020.03.011