Twiddler's syndrome with occult lead insulation damage in a child with epicardial pacemaker leads
We present a rare case of incidentally diagnosed Twiddler's syndrome in a child 7 years after implantation of a dual-chamber pacemaker system with epicardial leads. During revision, an insulation defect of the ventricular lead was evident, despite unremarkable prior pacemaker lead testing. The...
Gespeichert in:
Veröffentlicht in: | Cardiology in the young 2020-06, Vol.30 (6), p.890-891 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present a rare case of incidentally diagnosed Twiddler's syndrome in a child 7 years after implantation of a dual-chamber pacemaker system with epicardial leads. During revision, an insulation defect of the ventricular lead was evident, despite unremarkable prior pacemaker lead testing. The lead was repaired and a new generator was suture-fixated to prevent re-occurrence of generator manipulation. |
---|---|
ISSN: | 1047-9511 1467-1107 |
DOI: | 10.1017/S1047951120001031 |