Twiddler's syndrome with occult lead insulation damage in a child with epicardial pacemaker leads

We present a rare case of incidentally diagnosed Twiddler's syndrome in a child 7 years after implantation of a dual-chamber pacemaker system with epicardial leads. During revision, an insulation defect of the ventricular lead was evident, despite unremarkable prior pacemaker lead testing. The...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cardiology in the young 2020-06, Vol.30 (6), p.890-891
Hauptverfasser: Kramer, Peter, Berger, Felix, Peters, Björn
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present a rare case of incidentally diagnosed Twiddler's syndrome in a child 7 years after implantation of a dual-chamber pacemaker system with epicardial leads. During revision, an insulation defect of the ventricular lead was evident, despite unremarkable prior pacemaker lead testing. The lead was repaired and a new generator was suture-fixated to prevent re-occurrence of generator manipulation.
ISSN:1047-9511
1467-1107
DOI:10.1017/S1047951120001031