EVALUATION OF RADIATION DOSE FOR PATIENTS UNDERGOING MAMMOGRAPHY IN QATAR

Abstract In the absence of information on radiation doses in mammography in the Gulf countries, this study was designed to assess patient dose in terms of entrance surface air kerma and average glandular dose (AGD) in three mammography units in Qatar that covers 21% of all mammography systems in the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation protection dosimetry 2020-07, Vol.189 (3), p.354-361
Hauptverfasser: AlNaemi, Huda, Aly, Antar, J Omar, Ahmed, AlObadli, Amal, Ciraj-Bjelac, Olivera, Kharita, Mohamad Hassan, Rehani, Madan M
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract In the absence of information on radiation doses in mammography in the Gulf countries, this study was designed to assess patient dose in terms of entrance surface air kerma and average glandular dose (AGD) in three mammography units in Qatar that covers 21% of all mammography systems in the country. The study of 150 patients involving 600 projections indicated that the average value of AGD in patients was 2.2 mGy for cranio-caudal and 2.5 mGy for mediolateral-oblique views, respectively. Dose assessment was also performed for polymethyl methacrylate phantoms of thicknesses, ranging from 20 to 80 mm. Comparing the patient dose values with several other publications in literature for full-field digital mammography, our values are typically higher, which can be likely attributed to the larger compressed breast thickness.
ISSN:0144-8420
1742-3406
DOI:10.1093/rpd/ncaa049