Identifying a Break in the Chain: An Analysis of Where Ureteroscope Damage Occurs in the Hospital Cycle
To assess the timing and variables associated with damage to flexible ureteroscopes (fURS) at our institution. Flexible ureteroscopy is an important modality in the treatment of benign and malignant conditions of the upper urinary tract. While the durability and versatility of fURS have improved con...
Gespeichert in:
Veröffentlicht in: | Urology (Ridgewood, N.J.) N.J.), 2020-07, Vol.141, p.39-44 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To assess the timing and variables associated with damage to flexible ureteroscopes (fURS) at our institution. Flexible ureteroscopy is an important modality in the treatment of benign and malignant conditions of the upper urinary tract. While the durability and versatility of fURS have improved considerably, repair costs remain high and time out of commission diminishes workflow. After purchasing new digital fURS, we studied how and when these instruments were being damaged.
Between September 1, 2017 and June 30, 2018, we performed leak testing on fURS both before and after use. We gathered intraoperative data related to the user, the surgical indication, and the associated tools used in all cases that employed a digital or fiber optic fURS. Categorical and continuous variables were analyzed to identify risk factors for intraoperative fURS damage.
During the study period, complete data was gathered for 281 cases. Twenty-two fURS failed leak testing indicating an overall leak failure rate of 7.8%. Of these, 15 failed leak testing preoperatively indicating nonoperative damage occurring sometime during transport, handling, or sterile processing. The other 7 failures occurred during the procedures. No intraoperative variables were significantly associated with failures.
Our institutional leak failure rate is 8% (22/281). The majority of these failures did not occur during surgery. Of the 7 that occurred during surgery, larger stone burden and higher wattage showed mild association. Ongoing evaluation will target minimizing fURS damage outside of the operating room. |
---|---|
ISSN: | 0090-4295 1527-9995 |
DOI: | 10.1016/j.urology.2020.02.030 |