Leadless pacemakers reduce risk of device-related infection: Review of the potential mechanisms

Pacemaker-related infections remain a constant concern due to increased risk of patient morbidity and mortality. Although transvenous pacemakers are expected to have an infection rate ranging from 0.77% to 2.08%, no cases of leadless pacemaker infection have been reported in clinical trials enrollin...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Heart rhythm 2020-08, Vol.17 (8), p.1393-1397
Hauptverfasser: El-Chami, Mikhael F., Bonner, Matt, Holbrook, Reece, Stromberg, Kurt, Mayotte, Jane, Molan, Amy, Sohail, M. Rizwan, Epstein, Laurence M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Pacemaker-related infections remain a constant concern due to increased risk of patient morbidity and mortality. Although transvenous pacemakers are expected to have an infection rate ranging from 0.77% to 2.08%, no cases of leadless pacemaker infection have been reported in clinical trials enrolling more than 3000 patients. Many potential reasons why leadless pacemakers may be resistant to infection include the absence of a subcutaneous pocket and leads, reduced skin and glove contact, size, location, and device material. This review summarizes the current state of evidence regarding the apparent infection resistance of leadless pacemakers. [Display omitted]
ISSN:1547-5271
1556-3871
DOI:10.1016/j.hrthm.2020.03.019