Biodegradation ZK50 magnesium alloy compression screws: Mechanical properties, biodegradable characteristics and implant test
Magnesium alloy implants have lower stress load and can be absorbed gradually, but their degradation rates are too fast generally. A magnesium alloy contained 5% Zn and 0.5% Zr (ZK50) which have lower degradation rate are designed to be applied to cannulated bone screw. An oxidation heat treatment o...
Gespeichert in:
Veröffentlicht in: | Journal of orthopaedic science : official journal of the Japanese Orthopaedic Association 2020-11, Vol.25 (6), p.1107-1115 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Magnesium alloy implants have lower stress load and can be absorbed gradually, but their degradation rates are too fast generally. A magnesium alloy contained 5% Zn and 0.5% Zr (ZK50) which have lower degradation rate are designed to be applied to cannulated bone screw.
An oxidation heat treatment of 380 °C for 2 h proceeds to modify the ZK50 Mg alloy (ZK50-H). The microstructure observation, degradation tests and Biocompatibility analysis are proceeded between ZK50 and ZK50-H. Finally, a mini-pig implantation test is proceeded to provide a reference of implant application for future pre-clinical evaluation.
The heat treatment can improve the mechanical properties. A passive ceramic layer formed after simulated body fluid (SBF) solution immersion can restrict the degradation effectively. The cytotoxicity test shows the initial biosafety of ZK50 Mg alloy. A mini-pig implantation test of bone screw has proceeded to confirm the advanced biocompatibility. The ZK50-H screws can maintain enough support at least 8 weeks which the fracture of bone can get curing. The excellent osteoinduction of ZK50-H has a positive effect to growth of new bones and help the mini-pig regain heal faster in 12 weeks.
This study shows ZK50-H Mg alloy screw is a feasible degradation implant and can be carried out the next-step clinical experiments. |
---|---|
ISSN: | 0949-2658 1436-2023 |
DOI: | 10.1016/j.jos.2020.01.018 |