Safety and Efficacy of Initializing the Control-IQ Artificial Pancreas System Based on Total Daily Insulin in Adolescents with Type 1 Diabetes
To assess the safety and efficacy of a simplified initialization for the Tandem t:slim X2 Control-IQ hybrid closed-loop system, using parameters based on total daily insulin ("MyTDI") in adolescents with type 1 diabetes under usual activity and during periods of increased exercise. Adolesc...
Gespeichert in:
Veröffentlicht in: | Diabetes technology & therapeutics 2020-08, Vol.22 (8), p.594-601 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To assess the safety and efficacy of a simplified initialization for the Tandem t:slim X2 Control-IQ hybrid closed-loop system, using parameters based on total daily insulin ("MyTDI") in adolescents with type 1 diabetes under usual activity and during periods of increased exercise.
Adolescents with type 1 diabetes 12-18 years of age used Control-IQ for 5 days at home using their usual parameters. Upon arrival at a 60-h ski camp, participants were randomized to either continue Control-IQ using their home settings or to reinitialize Control-IQ with MyTDI parameters. Control-IQ use continued for 5 days following camp. The effect of MyTDI on continuous glucose monitoring outcomes were analyzed using repeated measures analysis of variance (ANOVA): baseline, camp, and at home.
Twenty participants were enrolled and completed the study; two participants were excluded from the analysis due to absence from ski camp (1) and illness (1). Time in range was similar between both groups at home and camp. A tendency to higher time 40% tended to show greater time in the euglycemic range improvements between baseline and home than users with ratios |
---|---|